A study of electrical properties and detection performances of the semi-insulating (SI) InP based detectors is presented. Detectors with a top P+ layer and a Schottky back contact give the charge collection efficiency about 90% and an energy resolution 3.7% (FWHM) for 5.48 MeV α-particles at 250 K. Detection of X-rays (122 keV and 60 keV) photons in temperature region 220-250 K is demonstrated. Multiple peaking observed during detection of photons is discussed.
Semi-insulating InP particle detectors for X- and γ-ray detection / Pelfer, P.G., Dubecky, F., Fornari, R., Pikna, M., Krempasky, M., Gombia, E., Darmo, J., Mosca, R., Sekacova, M.. - 487:(1998), pp. 477-481. (MRS Fall Meeting Symposium Boston, MA, USA, 1997).
Semi-insulating InP particle detectors for X- and γ-ray detection
Fornari R.Writing – Original Draft Preparation
;
1998-01-01
Abstract
A study of electrical properties and detection performances of the semi-insulating (SI) InP based detectors is presented. Detectors with a top P+ layer and a Schottky back contact give the charge collection efficiency about 90% and an energy resolution 3.7% (FWHM) for 5.48 MeV α-particles at 250 K. Detection of X-rays (122 keV and 60 keV) photons in temperature region 220-250 K is demonstrated. Multiple peaking observed during detection of photons is discussed.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.