A novel oscillator circuit with automatic capacitance compensation (ACC) capability has been coupled with quartz-crystal-microbalance (QCM) sensors coated with quinoxaline- and pyrazine-bridged cavitands to detect aromatic and chlorinated compounds in water. With double-side immersed 10-MHz crystals coated with the quinoxaline cavitand, the detection of toluene in deionized water was possible down to concentrations of a few parts per million. The ACC oscillator advantageously provides the simultaneous measurement of the sensor resonant frequency, damping, and value of the compensated parallel capacitance. This enabled observing that the analyte sorption in the cavitand coating not only brings about a mass uptake but also an increase of losses and, apparently, a rise in the coating average permittivity.

In-Liquid Sensing of Chemical Compounds by QCM Sensors Couplet with High-Accuracy ACC Oscillator / M., Ferrari; V., Ferrari; D., Marioli; A., Taroni; Suman, Michele; Dalcanale, Enrico. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 55:3(2006), pp. 828-834. [10.1109/TIM.2006.873792]

In-Liquid Sensing of Chemical Compounds by QCM Sensors Couplet with High-Accuracy ACC Oscillator

SUMAN, Michele;DALCANALE, Enrico
2006-01-01

Abstract

A novel oscillator circuit with automatic capacitance compensation (ACC) capability has been coupled with quartz-crystal-microbalance (QCM) sensors coated with quinoxaline- and pyrazine-bridged cavitands to detect aromatic and chlorinated compounds in water. With double-side immersed 10-MHz crystals coated with the quinoxaline cavitand, the detection of toluene in deionized water was possible down to concentrations of a few parts per million. The ACC oscillator advantageously provides the simultaneous measurement of the sensor resonant frequency, damping, and value of the compensated parallel capacitance. This enabled observing that the analyte sorption in the cavitand coating not only brings about a mass uptake but also an increase of losses and, apparently, a rise in the coating average permittivity.
2006
In-Liquid Sensing of Chemical Compounds by QCM Sensors Couplet with High-Accuracy ACC Oscillator / M., Ferrari; V., Ferrari; D., Marioli; A., Taroni; Suman, Michele; Dalcanale, Enrico. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 55:3(2006), pp. 828-834. [10.1109/TIM.2006.873792]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11381/1485503
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 30
  • ???jsp.display-item.citation.isi??? 29
social impact