A simple and reliable targeted liquid chromatography-electrospray-tandem mass spectrometry (LC-MS/MS) method was developed and validated through the selection of two biomarker peptides for the identification and determination of bovine insulin like growth factor-1 (IGF-1) in milk samples. Two urea-based sample extraction procedures were tested. The validation results provided detection limits at the 1–5 ng IGF-1/mL level as a function of the milk matrix, precision ranged from 3 to 8% and the method accuracy in the different milk matrices was assured. Finally, IGF-1 was measured in milk samples obtained by treatment with eleven different technological processes: IGF-1 concentrations were spread over a wide range from 11.2 ± 0.3 ng/mL to 346 ± 8 ng/mL with a median of 57.0 ± 0.2 ng/mL. The highest amount of IGF-1 was found in fresh whole milk samples and no significant correlation was found between the total milk protein content and the IGF-1 concentration level.

Development and single laboratory validation of a targeted liquid chromatography-triple quadrupole mass spectrometry-based method for the determination of insulin like growth factor-1 in different types of milk samples / Remaggi, G.; Saleri, R.; Andrani, M.; Satolli, F.; Rodighiero, E.; Elviri, L.. - In: FOOD CHEMISTRY X. - ISSN 2590-1575. - 13:(2022), p. 100271.100271. [10.1016/j.fochx.2022.100271]

Development and single laboratory validation of a targeted liquid chromatography-triple quadrupole mass spectrometry-based method for the determination of insulin like growth factor-1 in different types of milk samples

Remaggi G.;Saleri R.;Andrani M.;Satolli F.;Elviri L.
2022-01-01

Abstract

A simple and reliable targeted liquid chromatography-electrospray-tandem mass spectrometry (LC-MS/MS) method was developed and validated through the selection of two biomarker peptides for the identification and determination of bovine insulin like growth factor-1 (IGF-1) in milk samples. Two urea-based sample extraction procedures were tested. The validation results provided detection limits at the 1–5 ng IGF-1/mL level as a function of the milk matrix, precision ranged from 3 to 8% and the method accuracy in the different milk matrices was assured. Finally, IGF-1 was measured in milk samples obtained by treatment with eleven different technological processes: IGF-1 concentrations were spread over a wide range from 11.2 ± 0.3 ng/mL to 346 ± 8 ng/mL with a median of 57.0 ± 0.2 ng/mL. The highest amount of IGF-1 was found in fresh whole milk samples and no significant correlation was found between the total milk protein content and the IGF-1 concentration level.
2022
Development and single laboratory validation of a targeted liquid chromatography-triple quadrupole mass spectrometry-based method for the determination of insulin like growth factor-1 in different types of milk samples / Remaggi, G.; Saleri, R.; Andrani, M.; Satolli, F.; Rodighiero, E.; Elviri, L.. - In: FOOD CHEMISTRY X. - ISSN 2590-1575. - 13:(2022), p. 100271.100271. [10.1016/j.fochx.2022.100271]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11381/2919355
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact