Chromosomal DNA measurements represent the most accurate methodology for the quantitative analysis of human karyotype. A modified FACS 440 signal collection is described here. The FSC signal voltage coming from the optical bench has been split in two equal parts by a T-connector placed at the preamplifier entrance. The side scatter channel that is not employed during chromosome analysis has been used for the half-FSC signal. This modification allowed the contemporary use of FSC signal both for gating and analysis leading to a better discrimination of background, and therefore to a more accurate definition of sorting windows.
Technical improvements for flow karyotyping by standard FACS 440 flow cytometer / Vitale, Marco; Zamai, L.; Antonucci, A.; DEL GRANDE, P.; Papa, S.. - In: CYTOTECHNOLOGY. - ISSN 0920-9069. - 5(3):(1991), pp. 195-199.
Technical improvements for flow karyotyping by standard FACS 440 flow cytometer.
VITALE, Marco;
1991-01-01
Abstract
Chromosomal DNA measurements represent the most accurate methodology for the quantitative analysis of human karyotype. A modified FACS 440 signal collection is described here. The FSC signal voltage coming from the optical bench has been split in two equal parts by a T-connector placed at the preamplifier entrance. The side scatter channel that is not employed during chromosome analysis has been used for the half-FSC signal. This modification allowed the contemporary use of FSC signal both for gating and analysis leading to a better discrimination of background, and therefore to a more accurate definition of sorting windows.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.