Abstract We assessed the effect of intra-coronary attenuation on diagnostic accuracy using 64-slice computed tomography coronary angiography (CT-CA). We enrolled 170 patients with suspected coronary artery disease who underwent conventional coronary angiography (CA) and 64-slice CT-CA (100 ml of Iomeprol 400 mg I/ml at 4 ml/s). The study population was divided into two groups (85 patients each based on median attenuation of 326 HU) based on mean arterial attenuation; group 1 with low attenuation and group 2 with high attenuation. Diagnostic accuracy for the detection of significant coronary artery stenosis was determined for both groups using CA as reference standard. Overall, 163 significant stenoses were detected in 1,030 assessable coronary artery segments in group 1 compared with 160 significant stenoses in 1,020 assessable segments in group 2. The average intra-coronary attenuation was significantly (P< 0.05) higher for group 2 (388±46 HU) compared with group 1 (291±33 HU). The corresponding sensitivity and specificity values for detection of significant coronary artery stenosis were higher for group 2 (96.3% and 97.6%, respectively) than for group 1 (82.8% and 93.2%, respectively) and were more marked in distal coronary segments than in proximal segments. Higher intra-coronary attenuation on CT-CA results in greater diagnostic accuracy for detection of coronary artery stenosis. K

INFLUENCE OF INTRA-CORONARY ENHANCEMENT ON DIAGNOSTIC ACCURACY WITH 64-SLICE CT CORONARY ANGIOGRAPHY / Filippo, Cademartiri; Coruzzi, Paolo. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - 18:(2008), pp. 576-583. [10.1007/s00330-007-0773-0]

INFLUENCE OF INTRA-CORONARY ENHANCEMENT ON DIAGNOSTIC ACCURACY WITH 64-SLICE CT CORONARY ANGIOGRAPHY

CORUZZI, Paolo
2008-01-01

Abstract

Abstract We assessed the effect of intra-coronary attenuation on diagnostic accuracy using 64-slice computed tomography coronary angiography (CT-CA). We enrolled 170 patients with suspected coronary artery disease who underwent conventional coronary angiography (CA) and 64-slice CT-CA (100 ml of Iomeprol 400 mg I/ml at 4 ml/s). The study population was divided into two groups (85 patients each based on median attenuation of 326 HU) based on mean arterial attenuation; group 1 with low attenuation and group 2 with high attenuation. Diagnostic accuracy for the detection of significant coronary artery stenosis was determined for both groups using CA as reference standard. Overall, 163 significant stenoses were detected in 1,030 assessable coronary artery segments in group 1 compared with 160 significant stenoses in 1,020 assessable segments in group 2. The average intra-coronary attenuation was significantly (P< 0.05) higher for group 2 (388±46 HU) compared with group 1 (291±33 HU). The corresponding sensitivity and specificity values for detection of significant coronary artery stenosis were higher for group 2 (96.3% and 97.6%, respectively) than for group 1 (82.8% and 93.2%, respectively) and were more marked in distal coronary segments than in proximal segments. Higher intra-coronary attenuation on CT-CA results in greater diagnostic accuracy for detection of coronary artery stenosis. K
2008
INFLUENCE OF INTRA-CORONARY ENHANCEMENT ON DIAGNOSTIC ACCURACY WITH 64-SLICE CT CORONARY ANGIOGRAPHY / Filippo, Cademartiri; Coruzzi, Paolo. - In: EUROPEAN RADIOLOGY. - ISSN 0938-7994. - 18:(2008), pp. 576-583. [10.1007/s00330-007-0773-0]
File in questo prodotto:
File Dimensione Formato  
eur radiol.pdf

non disponibili

Tipologia: Documento in Post-print
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 264 kB
Formato Adobe PDF
264 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11381/1865921
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 77
  • ???jsp.display-item.citation.isi??? 71
social impact