Micromachined gas-chromatographic (GC) components in combination with high sensitivity MOX detectors and novel pre-concentration materials were utilized to develop a miniaturized, portable, stand-alone sub-ppb level monitoring system for aromatic volatile organic compounds (VOC). Using metal oxide semiconductor (MOX) gas sensors as detectors in a GC system, the intrinsic analytic selectivity of GC is exploited in a simplified architecture with on-board generated carrier gas. The work reported here includes the study of innovative pre-concentration materials, the micromachining of GC components and devices, the fabrication and characterization of complete system prototypes, and the development of application-specific control electronics. The results of both laboratory trials and successful real-world cross-validation with a reference analyzer are described.

Real-time monitoring of sub-ppb concentrations of aromatic volatiles with a MEMS-enabled miniaturized gas-chromatograph / Zampolli, S.; Elmi, I.; Mancarella, F.; Betti, Paolo; Dalcanale, Enrico; Cardinali, G. C.; Severi, M.. - In: SENSORS AND ACTUATORS. B, CHEMICAL. - ISSN 0925-4005. - 141:(2009), pp. 322-328. [10.1016/j.snb.2009.06.021]

Real-time monitoring of sub-ppb concentrations of aromatic volatiles with a MEMS-enabled miniaturized gas-chromatograph

BETTI, Paolo;DALCANALE, Enrico;
2009-01-01

Abstract

Micromachined gas-chromatographic (GC) components in combination with high sensitivity MOX detectors and novel pre-concentration materials were utilized to develop a miniaturized, portable, stand-alone sub-ppb level monitoring system for aromatic volatile organic compounds (VOC). Using metal oxide semiconductor (MOX) gas sensors as detectors in a GC system, the intrinsic analytic selectivity of GC is exploited in a simplified architecture with on-board generated carrier gas. The work reported here includes the study of innovative pre-concentration materials, the micromachining of GC components and devices, the fabrication and characterization of complete system prototypes, and the development of application-specific control electronics. The results of both laboratory trials and successful real-world cross-validation with a reference analyzer are described.
2009
Real-time monitoring of sub-ppb concentrations of aromatic volatiles with a MEMS-enabled miniaturized gas-chromatograph / Zampolli, S.; Elmi, I.; Mancarella, F.; Betti, Paolo; Dalcanale, Enrico; Cardinali, G. C.; Severi, M.. - In: SENSORS AND ACTUATORS. B, CHEMICAL. - ISSN 0925-4005. - 141:(2009), pp. 322-328. [10.1016/j.snb.2009.06.021]
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11381/2279387
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 190
  • ???jsp.display-item.citation.isi??? 168
social impact